Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. New experiment, all lasers on. 
  2. Set laser delays to 0
  3. Set all scaling factors to 0.5 (inc. FSC)
  4. Acquire data with QF scripter automatically changing the QF intensity every second - 1000 events/second, gaussian pulse - standard settings (FSC set to 400V)
  5. Set FSC so that all H/W FSC signals are on scale. 
  6. Set events to record to 20k.
  7. Record data with PMT volts from 0-1000 (25V increments) - PM for app to do this unattended (vchanger.exe)
  8. Put data in sub folder labelled 'laser_on' and 'laser_off' and run R script (updated in 2022)

SIGNAL:NOISE

VOLTAGES - V1 VOLTAGES  voltages_2022_04.xlsx *note voltages have been adjusted to minimise spectral unmixing artefacts.

...