  • Scattered signals are not displayed as they are irrelevant. Quantiflash does not emit well <400nm hence BUV395 plot is N/A & in the NIR hence why these plots look odd.
  • Recorded .fcs files need to be in a subfolder called 'laser_on'.
  • With this data - an optimal range of voltages can be determined. Next update will provide for minimal voltage where S/N plateaus as well as numerical deviation from linearity of the detector. 
  • Height parameter used - matched manual voltration in testing

Protocol - 

  1. New experiment, all lasers on. 
  2. Set laser delays to 0
  3. Set all scaling factors to 0.5 (inc. FSC)
  4. Acquire data with QF scripter automatically changing the QF intensity every second - 1000 events/second, gaussian pulse - standard settings.
  5. Set FSC so that all H/W FSC signals are on scale. 
  6. Set events to record to 200k20k.
  7. Record data with PMT volts from 0-1000 (25V increments) - PM for app to do this unattended (vchanger. exe)
  8. Put data in sub folder labelled 'laser_on' and run R script.

...