CS&T bead lot: 93308 23353
Laser (nm) | Power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Dichroic Mirror | Laser order | rSDen | 2.5x rSDen |
---|---|---|---|---|---|---|---|---|
405 | 50 | 1 | ||||||
V450_50B | Pacific Blue, Cascade Blue, AF405, V450, BV421, 450, DAPI, eFluor450, SparkViolet423, SuperBright436 | 450/50 | NA | 23.5 | 58.75 | |||
V510_50A | CFP, AF430, AmCyan, Cascade Yellow, V500, BV510, PO,eFluor506, SparkViolet538 | 510/50 | 502LP | 22 | 55 | |||
488 | 20 | 2 | ||||||
SSC | SSC | 488/10 | NA | NA | NA | |||
B530_30E | FITC, GFP, AF488, YFP, 5-FAM, CFSE, BB515, ZsGreen, SYBR Green, BODIPY FL | 530/30 | 502LP | 22.1 | 55.25 | |||
B585_42D | PE,PI,DsRed,DiA | 585/42 | 556LP | 19.8 | 49.5 | |||
B670_LPB | PerCP,PI, PerCP-Cy5.5,PerCP-Cy5,PE-Cy5,7AAD, BB700 | 670LP | 655LP | 21.7 | 54.25 | |||
B780_60A | PE-Cy7, PE-Vio770 | 780/60 | 735LP | 21.8 | 54.8 | |||
633 | 20 | 3 | ||||||
R660_20C | APC, AF647, CF647, DiD, DRAQ5, eFluor660 | 660/20 | NA | 21.7 | 54.25 | |||
R780_60A | APC-Cy7, APC-H7, DiR, APC-FIRE, APC-eFluor780, APCAPCVio770 | 780/60 | 735LP | 21.8 | 54.5 |
Important Information
- Baseline in use is #80998. Ensure you save new or update existing application settings after the 12th November 2018.#23533.
- For assistance setting up an existing experiment or using application settings please book a time with flow staff.
- Default voltages are normally good for resolving dim populations. Rule of thumb guide to settings voltages include, ensuring the rSD of the unlabelled cells is at least 2.5x the value of the rSD of electronic noise of the detector (that is viewable from the instrument configuration sheet) that accompanies each cytometer. You need to record a small file in order to accurately determine the rSD of the unlabelled cells each time. Remember this should be the minimum voltage and not the voltage to use. CST voltages are a good voltage to use with the exception that if a population is off scale, then the voltage needs to be reduced.
CantoII - ICPMR
Baseline: 03 2020
Baseline reports:
Cytometer Baseline Report_ICPMRReporticpmr.pdf
CS&T bead lot: 93308 23353
Laser (nm) | Power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Dichroic Mirror | Laser order | rSDen | 2.5x rSDen |
---|---|---|---|---|---|---|---|---|
405 | 50 | 1 | ||||||
V450_50B | Pacific Blue, Cascade Blue, AF405, V450, BV421, DAPI, eFluor450, SparkViolet423, SuperBright436 | 450/50 | NA | 16 | 40 | |||
V510_50A | CFP, AF430, AmCyan, Cascade Yellow, V500, BV510, PO, eFluor506, SparkViolet538 | 510/50 | 502LP | 16.1 | 40.25 | |||
488 | 20 | 2 | ||||||
SSC | SSC | 488/10 | NA | N/A | ||||
B530_30E | FITC, GFP, AF488, YFP, 5-FAM, CFSE, BB515, ZsGreen, SYBR Green, BODIPY FL | 530/30 | 502LP |
| 21.3 | 53.25 | ||
B585_42D | PE,PI,DsRed,DiA | 585/42 | 556LP | 22.9 | 57.25 | |||
B670_LPB | PerCP,PI, PerCP-Cy5.5,PerCP-Cy5,PE-Cy5,7AAD, BB700 | 670LP | 655LP | 22.2 | 55.5 | |||
B780_60A | PE-Cy7, PE-Vio770 | 780/60 | 735LP | 21.1 | 52.75 | |||
633 | 20 | 3 | ||||||
R660_20C | APC, AF647, CF647, DiD, DRAQ5, eFluor660 | 660/20 | NA | 16.1 | 40.25 | |||
R780_60A | APC-Cy7, APC-H7, DiR, APC-FIRE, APC-eFluor780, APCAPCVio770 | 780/60 | 735LP | 15.6 | 39 |
Important Information
- Baseline in use is #80998. Ensure you save new or update existing application settings after the 5th November 2018.For assistance setting up an existing experiment or #23533.
- For assistance setting up an existing experiment or using application settings please book a time with flow staff.
- Default voltages are normally good for resolving dim populations. Rule of thumb guide to settings voltages include, ensuring the rSD of the unlabelled cells is at least 2.5x the value of the rSD of electronic noise of the detector (that is viewable from the instrument configuration sheet) that accompanies each cytometer. You need to record a small file in order to accurately determine the rSD of the unlabelled cells each time. Remember this should be the minimum voltage and not the voltage to use. CST voltages are a good voltage to use with the exception that if a population is off scale, then the voltage needs to be reduced.
CantoII - SSM
...
Canto - CMRI
Baseline:
Baseline reports:
LSRII - CantoCMRI_Cytometer Baseline Report.pdf
LSRII - Cytometer Baseline Report.csv
CS&T bead lot: 80998 93308
...
Symphony - SSM - December 2019
CD4 Spread Profiles - WIMR - A5.2 - CD4 Spread Profiles.pptx
BD Generated SSM - WIMR - A5.2 - SSM.xlsx
Westmead Cytometry SSM -
Symphony Stain Index
ARIAIII - Cell Sorter
Baseline:
Baseline reports:
70um Cytometer Baseline Report.pdf
85um Cytometer Baseline Report.pdf
100um Cytometer Baseline Report.pdf
130um Cytometer Baseline Report.pdf
CS&T bead lot: 80998
633
18
20.5
Laser (nm) | Power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Dichroic Mirror | Laser order | rSDen | 2.5x rSDen | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
488 | 5020 | 1 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
SSC | SSC | 488/10 | NA | N/A | N/A | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
B530_30B | Alexa Fluor 488 | FITC, GFP, AF488, YFP, 5-FAM, CFSE, BB515, ZsGreen, SYBR Green, BODIPY FL | 530/30 | 505LP502LP |
| 2422. | 78 | 61.75 | B710_50A | PerCP-Cy5.5, 7AAD, BB700 | 710/50 | 685LP | 24.5 | 61.25 | 405 | 50 | 2 | V440_40F | Pacific Blue, Cascade Blue, AF405, V450, BV421, Dylight405 | 440/40 | NA | 20.1 | 50.25 | V525_50E | CFP, AF430, AmCyan, Cascade Yellow, V500, BV510, PO | 525/50 | 505LP |
| 23.3 | 58.25 | V610_20D | Qdot 605, BV605 | 610/20 | 600LP | 35.1 | 87.75 | V660_20C | Qdot 655, BV650 | 660/20 | 630LP | 22.6 | 56.5 | V710_50B | Qdot 700, BV711 | 710/50 | 685LP | 28.4 | 71 | V780_60A | BV785 | 780/60 | 750LP | 22.6 | 56.5 | 355 | 20 | 3 | U379_28C | BUV395 | 379/28 | NA | 17.9 | 44.75 | U515_30B | BUV496, DAPI, Hoechst Blue | 515/30 | 450LP | 16.5 | 41.25 | U740_35A | BUV737, Hoechst Red | 740/35 | 690LP | 19.2 | 48 | 640 | 40 | 4 | R670_14C | APC, AF647, CF647, DiD, DRAQ5, SYTOX RED | 670/14 | NA | 14.2 | 35.5 | R730_45B | APC-R700, AF700, DRAQ5, FVS700 | 730/45 | 690LP | 13.7 | 34.25 | R780_60A | APC-Cy7, APC-H7, DiR, APC-FIRE, APC-eFluor780, APC | 780/60 | 750LP | 14.7 | 36.75 | 561 | 100 | 5 | Y586_15D | PE, DsRed, dTomato, mTangerine, mStrawberry | 586/15 | 570LP | 17.7 | 44.25 | Y610_20C | PE-Texas Red, mCherry, PE-CF594, AF568, PI, PE-Dazzle, 7AAD, AF594 | 610/20 | 600LP | 16.4 | 41 | Y710_50B | PE-Cy5.5, 7AAD | 710/50 | 685LP | 16.4 | 41 | Y780_60A | PE-Cy7, PE-Vio770, | 780/60 | 750LP | 19.1 | 47.75 |
Important Information
- Baseline in use is #80998. Ensure you save new or update existing application settings after the 31st October 2018.
- Default voltages are normally good for resolving dim populations. Rule of thumb guide to settings voltages include, ensuring the rSD of the unlabelled cells is at least 2.5x the value of the rSD of electronic noise of the detector (that is viewable from the instrument configuration sheet) that accompanies each cytometer. You need to record a small file in order to accurately determine the rSD of the unlabelled cells each time. Remember this should be the minimum voltage and not the voltage to use. CST voltages are a good voltage to use with the exception that if a population is off scale, then the voltage needs to be reduced.
- Filter changes on the Fortessa and LSRII can be done once users have been trained. The “Non-default signage” must be displayed when using a non-default filter in place of the default.
- For optimal detection of PE-CY5, PE-AF647 & mPlum a 670/30 filter is needed in Y710_50B with a 635LP.
- For optimal detection of PerCP a 675/50 could be used in B710_50A with a 635LP. There are other alternatives too.
- For optimal detection of Hoechst Blue & AF350 a 450/50 bandpass filter could be used in U515_30B, with a 425LP.
- For optimal detection of AF633 a 670/40 bandpass filter could be used in R670_14C.
Fortessa
Baseline:
Baseline reports:
Cytometer Baseline Report_fortessa.pdf
CS&T bead lot: 93308
...
488
...
50-100
default @ 50
...
SSC
...
SSC
...
488/10
...
N/A
...
B530_30B
...
FITC, GFP, AF488, YFP, 5-FAM, CFSE, BB515, ZsGreen
...
530/30
...
505LP
...
403
...
50
...
CFP, AF430, AmCyan, Cascade Yellow, V500, BV510, PO
...
525/50
...
505LP
...
...
Qdot 605, BV605
...
610/20
...
600LP
...
Qdot 655, BV650
...
660/20
...
630LP
...
Qdot 700, BV711
...
710/50
...
685LP
...
355
...
20
...
639
...
40
...
R730_45B
...
APC-R700,AF700, DRAQ5, FVS700
...
730/45
...
690LP
...
R780_60A
...
APC-Cy7, APC-H7, DiR, APC-FIRE, APC-eFluor780, APC
...
780/60
...
750LP
...
561
...
50
...
PE-Texas Red, mCherry, PE-CF594, AF568, PI, PE-Dazzle
...
Important Information
- Baseline in use is #80998. Ensure you save new or update existing application settings after the 31st October 2018.
- Default voltages are normally good for resolving dim populations. Rule of thumb guide to settings voltages include, ensuring the rSD of the unlabelled cells is at least 2.5x the value of the rSD of electronic noise of the detector (that is viewable from the instrument configuration sheet) that accompanies each cytometer. You need to record a small file in order to accurately determine the rSD of the unlabelled cells each time. Remember this should be the minimum voltage and not the voltage to use. CST voltages are a good voltage to use with the exception that if a population is off scale, then the voltage needs to be reduced.
- Filter changes on the Fortessa and LSRII can be done once users have been trained. The “Non-default signage” must be displayed when using a non-default filter in place of the default.
- For optimal detection of PE-CY5, PE-AF647 & mPlum a 670/30 filter is needed in Y710_50B with a 635LP.
- For optimal detection of PerCP a 675/50 could be used in B710_50A with a 635LP. There are other alternatives too.
- For optimal detection of Hoechst Blue & AF350 a 450/50 bandpass filter could be used in U515_30B, with a 425LP.
- For optimal detection of AF633 a 670/40 bandpass filter could be used in R670_14C.
Symphony
Baseline: 20191127 - TBU
Baseline report: TBU
CS&T bead lot: 80998
...
594nm
Y-58 Schott
...
57 | ||||||||
PE | PE,PI,DsRed,DiA | 585/42 | 556LP | 22.8 | 57 | |||
PerCP | PerCP,PI, PerCP-Cy5.5,PerCP-Cy5,PE-Cy5,7AAD, BB700 | 670LP | 655LP | 22.8 | 57 | |||
PE-Cy7 | PE-Cy7, PE-Vio770 | 780/60 | 735LP | 20.5 | 51.25 | |||
633 | 20 | 2 | ||||||
APC | APC, AF647, CF647, DiD, DRAQ5, eFluor660 | 660/20 | NA | 22.8 | 57 | |||
APC-Cy7 | APC-Cy7, APC-H7, DiR, APC-FIRE, APC-eFluor780, APCVio770 | 780/60 | 735LP | 23.3 | 58.25 |
LSRII
Baseline:
Baseline reports:
Cytometer Baseline ReportLSRII.pdf
CS&T bead lot: 23353
Laser (nm) | Power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Dichroic Mirror | Laser order | rSDen | 2.5x rSDen |
---|---|---|---|---|---|---|---|---|
488 | 50 | 1 | ||||||
SSC | SSC | 488/10 | NA | N/A | N/A | |||
B530_30B | FITC, GFP, AF488, YFP, 5-FAM, CFSE, BB515, ZsGreen, SYBR Green, BODIPY FL | 530/30 | 505LP | 24.7 | 61.75 | |||
B710_50A | PerCP-Cy5.5, 7AAD, BB700, PerCP-eFluor710 | 710/50 | 685LP | 24.5 | 61.25 | |||
|
| |||||||
|
|
|
| |||||
|
|
| ||||||
|
|
| ||||||
|
|
| ||||||
355 | 20 | 3 | ||||||
U379_28C | BUV395 | 379/28 | NA | 17.9 | 44.75 | |||
U515_30B | BUV496, DAPI, Hoechst Blue | 515/30 | 450LP | 16.5 | 41.25 | |||
U740_35A | BUV737, Hoechst Red | 740/35 | 690LP | 19.2 | 48 | |||
640 | 40 | 4 | ||||||
R670_14C | APC, AF647, CF647, DiD, DRAQ5, SYTOX RED, eFluor660, SparkNIR685 | 670/14 | NA | 14.2 | 35.5 | |||
R730_45B | APC-R700, AF700, DRAQ5, FVS700 | 730/45 | 690LP | 13.7 | 34.25 | |||
R780_60A | APC-Cy7, APC-H7, DiR, APC-FIRE, APC-eFluor780, APCVio770 | 780/60 | 750LP | 14.7 | 36.75 | |||
561 | 100 | 5 | ||||||
Y586_15D | PE, DsRed, dTomato, mTangerine, mStrawberry, SparkYG581 | 586/15 | 570LP | 17.7 | 44.25 | |||
Y610_20C | PE-Texas Red, mCherry, PE-CF594, AF568, PI, PE-Dazzle, 7AAD, AF594, PE-eFluor610, SparkYG593 | 610/20 | 600LP | 16.4 | 41 | |||
Y710_50B | PE-Cy5.5, 7AAD, PE-Fire700 | 710/50 | 685LP | 16.4 | 41 | |||
Y780_60A | PE-Cy7, PE-Vio770, | 780/60 | 750LP | 19.1 | 47.75 |
Important Information
- Baseline in use is #23533. Violet laser is not functional at the moment.
- Default voltages are normally good for resolving dim populations. Rule of thumb guide to settings voltages include, ensuring the rSD of the unlabelled cells is at least 2.5x the value of the rSD of electronic noise of the detector (that is viewable from the instrument configuration sheet) that accompanies each cytometer. You need to record a small file in order to accurately determine the rSD of the unlabelled cells each time. Remember this should be the minimum voltage and not the voltage to use. CST voltages are a good voltage to use with the exception that if a population is off scale, then the voltage needs to be reduced.
- Filter changes on the Fortessa and LSRII can be done once users have been trained. The “Non-default signage” must be displayed when using a non-default filter in place of the default.
- For optimal detection of PE-CY5, PE-AF647 & mPlum a 670/30 filter is needed in Y710_50B with a 635LP.
- For optimal detection of PerCP a 675/50 could be used in B710_50A with a 635LP. There are other alternatives too.
- For optimal detection of Hoechst Blue & AF350 a 450/50 bandpass filter could be used in U515_30B, with a 425LP.
- For optimal detection of AF633 a 670/40 bandpass filter could be used in R670_14C.
Fortessa
Baseline target value reset:
Baseline reports:
Reset Target Values ReportFortessa.pdf
Cytometer Baseline Report_fortessa.pdf
CS&T bead lot: 23353
Laser (nm) | Power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Dichroic Mirror | Laser order | rSDen | 2.5x rSDen |
---|---|---|---|---|---|---|---|---|
488 | 50-100default @ 50 | 1 | ||||||
SSC | SSC | 488/10 | N/A | N/A | N/A | |||
B530_30B | FITC, GFP, AF488, YFP, 5-FAM, CFSE, BB515, ZsGreen, BODIPY FL, CF488A, DyLight488, EGFP, EYFP, Kiravia Blue520, NovaFluor Blue510, NovaFluor Blue530, SparkBlue550, Sytox Green, VioBright515, VioBright FITC | 530/30 | 505LP | 15.1 | 37.75 | |||
B710_50A | PerCP-Cy5.5, 7AAD, BB700, PerCP-eFluor710, PE-Fire700, PerCP-Vio700 | 710/50 | 685LP | 15.1 | 37.75 | |||
403 | 50 | 2 | ||||||
V440_40F | Pacific Blue, Cascade Blue, AF405, V450, BV421, Dylight405, DAPI, eFluor450, SparkViolet423, SuperBright436, CellTrace Violet, CF405S, StarBright V440, Sytox Blue, VioBlue | 440/40 | N/A | 10.1 | 25.25 | |||
V525_50E | Pacific Green, CFP, AF430, AmCyan, Cascade Yellow, V500, BV510, PO, eFluor506, SparkViolet538, Krome Orange, StarBright V515, T-Sapphire, VioGreen, VioletFluor500 | 525/50 | 505LP |
| 10.1 | 25.25 | ||
V610_20D | Qdot 605, BV605, SuperBright600, StarBright V610 | 610/20 | 600LP | 10.5 | 26.25 | |||
V660_20C | Qdot 655, BV650, SuperBright645, StarBright V670 | 660/20 | 630LP | 11.4 | 28.5 | |||
V710_50B | Qdot 700, BV711, SuperBright702, StarBright V710 | 710/50 | 685LP | 10.5 | 26.25 | |||
V780_60A | BV785, SuperBright780, StarBright V760, StarBright V790 | 780/60 | 750LP | 10 | 25 | |||
355 | 20 | 3 | ||||||
U379_28C | BUV395, StarBright UV400 | 379/28 | N/A | 20.5 | 51.25 | |||
U515_30B | BUV496, DAPI, Hoechst Blue, StarBright UV510 | 515/30 | 450LP | 22.1 | 55.25 | |||
U740_35A | BUV737, Hoechst Red, StarBright UV740 | 740_35 | 690LP | 20.4 | 51 | |||
639 | 40 | 4 | ||||||
R670_14C | APC, AF647, CF647, DiD, DRAQ5, eFluor660, SparkNIR685, Celltrace Far Red, DyLight633, DyLight649, NovaFluor Red660, NovaFluor Red685, Sytox ORed | 670/14 | N/A | 15.7 | 39.25 | |||
R730_45B | APC-R700, AF680, AF700, DRAQ5, FVS700, NovaFluor Red700, R718 | 730/45 | 690LP | 16.4 | 41 | |||
R780_60A | APC-Cy7, APC-H7, DiR, APC-Fire750, APC-eFluor780, APCVio770, AF750, APC-AF750 | 780/60 | 750LP | 17 | 42.5 | |||
561 | 50 | 5 | ||||||
Y586_15D | PE, DsRed, dTomato, mTangerine, mOrange, mStrawberry,SparkYG581, CellTrace Yellow, DyLight549, eFluor570, NovaFluor Yellow570, RY586, Sytox Orange | 586/15 | 570LP | 13.6 | 34 | |||
Y610_20C | PE-Texas Red, mCherry, PE-CF594, AF568, AF594, PI, PE-Dazzle594, PE eFluor610, SparkYG593, BODPIY 581-591-C11, eFluor615, NovaFluor Yellow610, PE-Vio615, StarBright Y605 | 610/20 | 600LP | 14.9 | 37.25 | |||
Y710_50B | PE-Cy5.5, 7AAD, PE-Fire700, NovaFluor Yellow690, NovaFluor Yellow700, StarBright Y720 | 710/50 | 685LP | 13.6 | 34 | |||
Y780_60A | PE-Cy7, PE-Vio770, | 780/60 | 750LP | 15.5 | 38.75 |
Important Information
- Baseline in use is #23533.
- Default voltages are normally good for resolving dim populations. Rule of thumb guide to settings voltages include, ensuring the rSD of the unlabelled cells is at least 2.5x the value of the rSD of electronic noise of the detector (that is viewable from the instrument configuration sheet) that accompanies each cytometer. You need to record a small file in order to accurately determine the rSD of the unlabelled cells each time. Remember this should be the minimum voltage and not the voltage to use. CST voltages are a good voltage to use with the exception that if a population is off scale, then the voltage needs to be reduced.
- Filter changes on the Fortessa and LSRII can be done once users have been trained. The “Non-default signage” must be displayed when using a non-default filter in place of the default.
- For optimal detection of PE-CY5, PE-AF647 & mPlum a 670/30 filter is needed in Y710_50B with a 635LP.
- For optimal detection of PerCP a 675/50 could be used in B710_50A with a 635LP. There are other alternatives too.
- For optimal detection of Hoechst Blue & AF350 a 450/50 bandpass filter could be used in U515_30B, with a 425LP.
- For optimal detection of AF633 a 670/40 bandpass filter could be used in R670_14C.
Symphony
Baseline:
Baseline report:
Cytometer Baseline Reportsymphony.pdf
CS&T bead lot: 23353
Laser Colour | Laser (nm) | Max power (mw) | CST default power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Dichroic mirror | Notch filter | Laser order | Pinhole position | Detector specifics | Orientation | Additional notes |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Red | 637 | 140 | 140 | 1 | 1 | 5D | |||||||
R820_60 | APC-Cy7, APC-H7, DiR, APC-Vio770, APC-Fire810 | 820/60 BP | 800LP | H14121-50 | |||||||||
R780_60 | AF750, APC-AF750, APC-eFluor780, APC-Fire750, APC-H7 | 780/60 BP | 750 LP | 785nm | H14121-50 | ||||||||
R730_45 | APC-R700, AF700, DRAQ5, FVS700, AF680, NovaFluor Red685, NovaFluor Red700, R718 | 730/45 BP | 685 LP | H11900-20 | |||||||||
R670_30 | APC, AF647, CF647, DiD, DRAQ5, eFluor660, Spark NIR685, CellTrace Far Red, DyLight633, DyLight649, NovaFluor Red660, Sytox Red | 670/30 BP | 665 LP | H11900-20 | Schott RG665 | ||||||||
UV | 355 | 100 | 100 | 2 | 3 | M1&M2&M1 | |||||||
U820_60 | BUV805 | 820/60 BP | 800 LP | 785nm | H14121-50 | ||||||||
U780_60 | spectral | 780/60 BP | 770 LP | H14121-50 | |||||||||
U740_35 | BUV737, Hoechst Red, StarBright UV740 | 740/35 BP | 710 LP | H14121-50 | |||||||||
U695_40 | 695/40 | 685 LP | |||||||||||
U670_30 | BUV661, StarBright UV665 | 670/30 BP | 635 LP | H11900-20 | |||||||||
U610_20 | BUV615, StarBright UV610, StarBright UV605 | 610/20 BP | 600 LP | H11900-20 | |||||||||
U586_15 | BUV563 | 586/15 BP | 570 LP | 594nm | H11900-20 | ||||||||
U515_30 | BUV496, DAPI, Hoechst Blue, StarBright UV510 | 515/30 BP | 450 LP | H11900-20 | |||||||||
U450_50 | AF350, DAPI, CellTrace Blue, StarBright UV445 | 450/50 BP | 410 LP | H11900-20 | |||||||||
U379_28 | BUV395, StarBright UV400 | 379/28 BP | H11900-110 | ||||||||||
Yellow | 561 | 150 | 150 | 3 | 4 | 6D | |||||||
Y820_60 | PE-Cy7, StarBright Y800 | 820/60 BP | 800 LP | H14121-50 | |||||||||
Y780_60 | 780/60 BP | 750 LP | 785nm | H11900-20 | |||||||||
Y710_50 | PE-Cy5.5, PE-Fire700, NovaFluor Yellow690, NovaFluor Yellow700, StarBright Y720 | 710/50 BP | 685 LP | H11900-20 | |||||||||
Y670_30 | PE-Cy5, NovaFluor Yellow660, StarBright Y665 | 670/30 BP | 635 LP | H11900-20 | |||||||||
Y610_20 | PE-CF594, PE-Texas Red, mCherry, AF568, AF594, PI, PE-Dazzle594, PE eFluor610, SparkYG593, NovaFluor Yellow610, PE-Vio615, StarBright Y605 | 610/20 BP | 600 LP | H11900-20 | |||||||||
Y586_15 | PE, DsRed, dTomato, mTangerine, mOrange, mStrawberry, Spark YG581, CellTrace Yellow, DyLight549, eFluor570, NovaFluor Yellow570, R586, Sytox Orange | 586/15 BP | 594nm Y-58 Schott | H11900-40 | |||||||||
Blue | 488 | 200 | 150 | 4 | 5 | M1&M2 | |||||||
B820_60 | BB790, StarBright Blue810 | 820/60 BP | 800 LP | H14121-50 | |||||||||
B780_60 | RB780 | 780/60 BP | 750 LP | 785nm | H14121-50 | ||||||||
B750_30 | BB755, StarBright Blue 765 | 750/30 BP | 735 LP | H14121-50 | |||||||||
B710_50 | BB700, PerCP-Cy5.5, 7AAD, PerCP-eFluor710, PerCP-Vio700, PE-Fire700, StarBright Blue700 | 710/50 BP | 685 LP | H11900-20 | |||||||||
B670_30 | BB660, PerCP, NovaFluor Blue660-40S, NovaFluor Blue610-120S, StarBright Blue675 | 670/30 BP | 635 LP | H11900-20 | |||||||||
B610_20 | BB630, NovaFluor Blue610-30S, NovaFluor Blue610-70S, StarBright Blue615 | 610/20 BP | 600 LP | 594nm | H11900-20 | ||||||||
B515_30 | FITC, BB515, A488, BODIPY FL, CF488A, CSFE, DyLight488, EGFP, Kiravia Blue510, NF-Blue510, NF-Blue530, Spark Blue550, Sytox Green, VioBright515, VioBright FITC | 515/30 BP | 500 LP | H11900-40 | |||||||||
SSC | 488/10 BP | H11900-20 | |||||||||||
Orange | 594 | 100 | 100 | 5 | 6 | 5D | |||||||
O660_20 | spectral | 660/20 BP | 635 LP | H14121-20 | |||||||||
O616_23 | AF594, mCherry | 616/23 BP | 610 LP | H14121-20 | |||||||||
Violet | 406 | 200 | 200 | 6 | 7 | M1&M2&M1 | |||||||
V820_60 | BV786, BV875, SuperBright780, StarBright V790 | 820/60 BP | 800 LP | H14121-50 | |||||||||
V780_60 | spectral | 780/60 BP | 770 LP | 785nm | H14121-50 | ||||||||
V750_30 | BV750, StarBright V760 | 750/30 BP | 735 LP | H14121-50 | |||||||||
V710_50 | BV711, Qdot700, SuperBright702, StarBright V710 | 710/50 BP | 685 LP | H11900-20 | |||||||||
V677_20 | BV650, Qdot650, SuperBright645, StarBright V670 | 677/20 BP | 635 LP | H11900-20 | |||||||||
V610_20 | BV605, Qdot605, SuperBright600, StarBright V610 | 610/20 BP | 600 LP | H11900-20 | |||||||||
V586_15 | BV570, Pacific Orange | 586/15 BP | 550 LP | 594nm | H11900-20 | ||||||||
V525_50 | BV510, Pacific Green, CFP, AF430, AmCyan, Cascade Yellow, V500, PO, eFluor506, Spark Violet538, StarBright V515, T-Sapphire, VioGreen, VioletFluor500 | 525/50 BP | 505 LP | H11900-20 | |||||||||
V474_25 | BV480, DAPI, StarBright V475 | 474/25 BP | 450 LP | H11900-20 | |||||||||
V427_25 | BV421, Pacific Blue, Cascade Blue, AF405, V450, eFluor450, VioBlue, Spark Violet423, StarBright V440, SuperBright436, Sytox Blue | 427/25 BP | 410 LP | H11900-20 | |||||||||
V405_10 | Violet SSC | 405/10 BP | H11900-20 | ||||||||||
*PMT installed | H11900-20 | ||||||||||||
*PMT installed | H11900-20 | ||||||||||||
*PMT installed | H11900-20 | ||||||||||||
IndigoNIR | 446_779 | 75_100 | 75_100 | 7 | 9 | 5D | |||||||
N820_60 | DyLight800 (strep-avidin - recommended), AF790 | 820/60 BP | 800 LP | H14121-50 | |||||||||
I660_20 | spectral | 660/20 BP | 635 LP | H14121-20 | |||||||||
I605_40 | spectral | 605/40 BP | 600 LP | H11900-20 | |||||||||
I515_30 | spectral | 515/30 BP | 500 LP | H11900-20 | |||||||||
I470_15 | CFP, Chromomycin | 470/15 BP | 450 LP | H11900-20 |
Symphony - SSM - December 2020
2019 - BD - generated - CD4 Spread Profiles - WIMR - A5.2 - CD4 Spread Profiles.pptx
2019 - BD - generated - SSM - WIMR - A5.2 - SSM.xlsx
2020 - Westmead Cytometry SSM - 2020_SSM_symphony.xlsx
2019 - BD - generated - Symphony Stain Index
Aurora
Baseline target value reset:
Baseline reports:
CS&T bead lot: 200
Laser Colour | Laser (nm) | Max power (mw) | CST default power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Wavelength Start (nm) | Wavelength End (nm) | Laser order |
---|---|---|---|---|---|---|---|---|---|
Ultraviolet | 355 | 1 | |||||||
UV1 | 373/15 | 365 | 380 | ||||||
UV2 | 388/15 | 380 | 395 | ||||||
UV3 | 428/15 | 420 | 435 | ||||||
UV4 | 443/15 | 436 | 451 | ||||||
UV5 | 458/15 | 451 | 466 | ||||||
UV6 | 473/15 | 466 | 481 | ||||||
UV7 | 514/28 | 500 | 528 | ||||||
UV8 | 542/28 | 528 | 566 | ||||||
UV9 | 582/31 | 566 | 597 | ||||||
UV10 | 613/31 | 597 | 628 | ||||||
UV11 | 664/27 | 651 | 678 | ||||||
UV12 | 692/28 | 678 | 706 | ||||||
UV13 | 720/29 | 706 | 735 | ||||||
UV14 | 750/30 | 735 | 765 | ||||||
UV15 | 780/30 | 765 | 795 | ||||||
UV16 | 812/34 | 795 | 829 | ||||||
Violet | 405 | 2 | |||||||
V1 | 428/15 | 420 | 435 | ||||||
V2 | 443/15 | 436 | 451 | ||||||
V3 | 458/15 | 451 | 466 | ||||||
V4 | 473/15 | 466 | 481 | ||||||
V5 | 508/20 | 498 | 518 | ||||||
V6 | 525/17 | 516 | 533 | ||||||
V7 | 542/17 | 533 | 550 | ||||||
V8 | 581/19 | 571 | 590 | ||||||
V9 | 598/20 | 588 | 608 | ||||||
V10 | 615/20 | 605 | 625 | ||||||
V11 | 664/27 | 651 | 678 | ||||||
V12 | 692/28 | 678 | 706 | ||||||
V13 | 720/29 | 706 | 735 | ||||||
V14 | 750/30 | 735 | 765 | ||||||
V15 | 780/30 | 765 | 795 | ||||||
V16 | 812/34 | 795 | 829 | ||||||
Blue | 488 | 3 | |||||||
B1 | 508/20 | 498 | 518 | ||||||
B2 | 525/17 | 516 | 533 | ||||||
B3 | 542/17 | 533 | 550 | ||||||
B4 | 581/19 | 571 | 590 | ||||||
B5 | 589/20 | 588 | 608 | ||||||
B6 | 615/20 | 605 | 625 | ||||||
B7 | 661/17 | 654 | 670 | ||||||
B8 | 679/18 | 670 | 688 | ||||||
B9 | 697/19 | 688 | 707 | ||||||
B10 | 717/20 | 707 | 727 | ||||||
B11 | 738/31 | 728 | 749 | ||||||
B12 | 760/23 | 749 | 772 | ||||||
B13 | 783/23 | 772 | 795 | ||||||
B14 | 812/34 | 795 | 829 | ||||||
Yellow-Green | 561 | 4 | |||||||
YG1 | 577/20 | 567 | 587 | ||||||
YG2 | 598/20 | 588 | 608 | ||||||
YG3 | 615/20 | 605 | 625 | ||||||
YG4 | 661/17 | 653 | 670 | ||||||
YG5 | 679/18 | 670 | 688 | ||||||
YG6 | 697/19 | 688 | 707 | ||||||
YG7 | 720/29 | 706 | 735 | ||||||
YG8 | 750/30 | 735 | 765 | ||||||
YG9 | 780/30 | 765 | 795 | ||||||
YG10 | 812/34 | 795 | 829 | ||||||
Red | 640 | 5 | |||||||
R1 | 661/17 | 653 | 670 | ||||||
R2 | 679/18 | 670 | 688 | ||||||
R3 | 697/19 | 688 | 707 | ||||||
R4 | 717/20 | 707 | 727 | ||||||
R5 | 738/21 | 728 | 749 | ||||||
R6 | 760/23 | 749 | 772 | ||||||
R7 | 783/23 | 772 | 795 | ||||||
R8 | 812/34 | 795 | 829 |
ARIAIII - Cell Sorter
Baseline:
Baseline reports:
70umCytometer Baseline Report.pdf
85umCytometer Baseline Report.pdf
100um Cytometer Baseline Report.pdf
130umCytometer Baseline Report.pdf
CS&T bead lot: 23533
Laser (nm) | Power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Dichroic Mirror | Laser order | rSDen (70um) | 2.5x rSDen (70um) |
---|---|---|---|---|---|---|---|---|
633 | 18 | 1 | ||||||
R660_20C | APC, AF647, CF647, DiD, DRAQ5, eFluor660 | 660/20 | NA | 7.8 | 19.5 | |||
R730_45B | APC-R700, AF700, DRAQ5 | 730/45 | 690LP | 6.7 | 16.75 | |||
R780_60A | APC-Cy7, APC-H7, DiR, APC-eFL780, APC-Fire750 | 780/60 | 750LP | 6.1 | 15.25 | |||
488 | 20 | 2 | ||||||
SSC | 488/10 | NA | N/A | |||||
B530_30E | FITC, GFP, AF488, BBYFP, 5-FAM, CFSE , BB515, ZsGreen, BODIPY FL | 530/30 | 502LP | 7.1 | 17.75 | |||
B585_42D | PE, DsRed, dTomato (sub-optimal), mTangerine (sub-optimal), mStrawberry (sub-optimal) | 585/42 | 556LP | 7 | 17.5 | |||
B616_23C | PE-Texas Red, mCherry (sub-optimal), PE-CF594, PE-AF610, AF568, PE-eFL610, PE-Dazzle594, PE eFluor610 | 616/23 | 610LP | 7 | 17.5 | |||
B695_40B | PerCP, PerCP-Cy5.5, 7AAD, PI, PerCP-eFL710, PerCP-Vio700, BB700, PerCP-eFluor710 | 695/40 | 655LP | 7.6 | 19 | |||
B780_60A | PE-Cy7, PE-Vio770 | 780/60 | 735LP | 7 | 17.5 | |||
405 | 50 | 3 | ||||||
V440_40F | Pacific Blue, Cascade Blue, AF405, V450, DAPI, BV421, DyLight 405, eFL450, VioBlue, DAPI, eFluor450, SuperBright436 | 440/40 | NA | 6.2 | 15.5 | |||
V525_50E | CFP, AF430, AmCyan, Cascade Yellow, V500, BV510, PO, eFluor506 | 525/50 | 505LP | 5.1 | 12.75 | |||
V610_20D | BV605, eFL605, Qdot 605, SuperBright600 | 610/20 | 600LP | 5 | 12.5 | |||
V660_20C | BV650, eFL650, Qdot 655 | 660/20 | 630LP | 5.1 | 12.75 | |||
V710_50B | BV711, Qdot 700, SuperBright702 | 710/50 | 685LP | 6 | 15 | |||
V780_60A | BV785, Qdot 800, SuperBright780 | 780/60 | 750LP | 7.1 | 17.75 |
INFLUX - Cell Sorter
Filter configuration:
View file | ||||
---|---|---|---|---|
|
Laser (nm) | Power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Dichroic Mirror | Laser order |
---|---|---|---|---|---|---|
488 | 50-200default 100 | 1 | ||||
PMT 1 - FSC (Par) | Forward Scatter - Parrallel | 488/10 | NA | |||
PMT 2 - FSC (Perp) | Forward Scatter - Perpendicular | 488/10 | NA | |||
PMT 3 - SSC | Side Scatter | 488/10 | NA | |||
PMT 4 | FITC, GFP, AF488, BBYFP, 5-FAM, CFSE , BB515, ZsGreen, DiO, BODIPY FL | 530/30 | 600LP | |||
PMT 5 | PerCP, PerCP-Cy5.5, 7AAD, PI, PerCP-eFL710, PerCP-Vio700, BB700, PerCP-eFluor710 | 710/50 | NA | |||
405 | 100 | 405 nm - 100mw | 2 | |||
PMT 14 | Pacific Blue, Cascade Blue, AF405, V450, BV421, DyLight 405, eFL450, VioBlue, DAPI, eFluor450 | 440/40 | 455LP, 500LP | |||
PMT 15 | CFP, eCFP, eFluor506 | 480/40 | 500LP | |||
PMT 16 | CFP, AF430, AmCyan, Cascade Yellow, V500, BV510, PO, VioGreen | 525/30 | 550LP | |||
PMT 17 | BV570, Qdot 565, | 572/28 | 585LP, 645LP | |||
PMT 18 | BV605, DyLight405, Qdot 605, SuperBright600 | 610/20 | 645LP | |||
PMT 19 | BV650, eFL650, Qdot 655, SuperBright645 | 660/20 | 685LP | |||
PMT 20 | BV711, Qdot 700, SuperBright702 | 710/50 | NA | |||
561 | 150 | 561 nm - 150mw | 3 | |||
PMT 6 | PE, DsRed, dTomato, mTangerine, mStrawberry, Dil | 585/29 | 600LP | |||
PMT 7 | PE-Texas Red, mCherry, PE-CF594, PE-AF610, AF568, PE-eFL610, PE-Dazzle594, PE eFluor610 | 610/20 | 645LP | |||
PMT 8 | PE-Cy5 | 670/30 | 685LP | |||
PMT 9 | PE-Cy5.5 | 710/50 | 740LP | |||
PMT 10 | PE-Cy7, PE-Vio770 | 750LP | NA | |||
355 | 100 | 355 nm - 100mw | 4 | |||
PMT 13 | DAPI, Hoechst Blue | 460/50 | 480LP, 600LP | |||
PMT 22 | BUV395 | 379/34 | 427LP | |||
PMT 23 | BUV496 | 520/35 | 550LP, 600LP | |||
PMT 24 | BUV661, Hoechst Red | 670/30 | NA | |||
PMT_Unplugged | BUV563 | 585/42 | NA | |||
628 | 120 | 640 nm - 120mw | 5 | |||
PMT 11 | APC, AF647, CF647, DiD, DRAQ5,eFluor660 | 670/30 | 700LP | |||
PMT 12 | AF700, DRAQ5, APC-R700 | 740/40 | 740LP | |||
PMT 21 | APC-Cy7, APC-H7, DiR, APC-eFL780, APC-Fire750 |
488
20
21.25
21.5
23
25
405
50
18
16.5
INFLUX - Cell Sorter
Filter configuration:
View file | ||||
---|---|---|---|---|
|
Power
(mw)
488
50-200
default 100
750LP | NA |
BD FASDiscover S8 Imaging cell sorter
Laser Colour | Laser (nm) | Max power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Laser order |
---|---|---|---|---|---|---|
Imaging | 488 | 40 | FSC | N/A | ||
SSC (Imaging) | 488/15 | |||||
Light Loss (Imaging) | 488/5 | |||||
ImgB1 | 534/46 | |||||
ImgB2 | 600/60 | |||||
ImgB3 | 788/225 | |||||
Ultraviolet | 355 | 60 | 1 | |||
UV1 | 375/20 | |||||
UV2 | 392/14 | |||||
UV3 | 420/20 | |||||
UV4 | 440/20 | |||||
UV5 | 459/18 | |||||
UV6 | 476/18 | |||||
UV7 | 501/16 | |||||
UV8 | 517/16 | |||||
UV9 | 532/16 | |||||
UV10 | 547/16 | |||||
UV11 | 574/16 | |||||
UV12 | 590/16 | |||||
UV13 | 607/18 | |||||
UV14 | 625/18 | |||||
UV15 | 655/20 | |||||
UV16 | 675/20 | |||||
UV17 | 698/26 | |||||
UV18 | 724/26 | |||||
UV19 | 752/30 | |||||
UV20 | 782/30 | |||||
UV21 | 812/30 | |||||
UV22 | 844/24 | |||||
Violet | 405 | 100 | 2 | |||
SSC (Violet) | 405/15 | |||||
Lightloss (Violet) | 405/5 | |||||
V1 | 420/20 | |||||
V2 | 440/20 | |||||
V3 | 459/18 | |||||
V4 | 476/16 | |||||
V5 | 501/16 | |||||
V6 | 517/16 | |||||
V7 | 532/16 | |||||
V8 | 547/16 | |||||
V9 | 574/16 | |||||
V10 | 590/16 | |||||
V11 | 607/18 | |||||
V12 | 625/18 | |||||
V13 | 655/20 | |||||
V14 | 775/20 | |||||
V15 | 698/26 | |||||
V16 | 724/26 | |||||
V17 | 752/30 | |||||
V18 | 782/30 | |||||
V19 | 812/30 | |||||
V20 | 844/34 | |||||
Blue | 488 | 40 | 3 | |||
B1 | 501/16 | |||||
B2 | 517/16 | |||||
B3 | 532/16 | |||||
B4 | 547/16 | |||||
B5 | 574/16 | |||||
B6 | 590/16 | |||||
B7 | 607/18 | |||||
B8 | 625/18 | |||||
B9 | 655/20 | |||||
B10 | 675/20 | |||||
B11 | 698/26 | |||||
B12 | 724/30 | |||||
B13 | 752/30 | |||||
B14 | 782/30 | |||||
B15 | 812/30 | |||||
B16 | 844/34 | |||||
Yellow-Green | 561 | 100 | 4 | |||
YG1 | 574/16 | |||||
YG2 | 590/16 | |||||
YG3 | 607/18 | |||||
YG4 | 625/18 | |||||
YG5 | 655/20 | |||||
YG6 | 675/20 | |||||
YG7 | 698/26 | |||||
YG8 | 724/26 | |||||
YG9 | 752/30 | |||||
YG10 | 782/30 | |||||
YG11 | 812/30 | |||||
YG12 | 844/34 | |||||
Red | 640 | 100 | 5 | |||
R1 | 655/20 | |||||
R2 | 675/20 | |||||
R3 | 698/26 | |||||
R4 | 724/26 | |||||
R5 | 752/30 | |||||
R6 | 782/30 | |||||
R7 | 812/30 | |||||
R8 | 844/34 |
Cytoflex - WIMR
Baseline: NA
QC bead lot #: 12APA
Laser (nm) | Power (mw) | Detector | Common fluorochromes detectable (not exhaustive) | Band pass filter | Dichroic Mirror | Laser order |
---|---|---|---|---|---|---|
405 | 50 | 2 | ||||
V405_10 | vSSC | 405/10 | NA | |||
V450_50 | DAPI, BV421, Pacific Blue, Cascade Blue, AF405, V450, DyLight 405, eFL450, VioBlue | 450/45 | NA | |||
488 | 20 | 1 | ||||
SSC | SSC | 488/8 | NA | |||
B525_40 | FITC, GFP, AF488, YFP, 5-FAM, CFSE, BB515, ZsGreen, SYBR Green | 525/40 | NA | |||
B585_42 | PE,PI,DsRed,DiA | 585/42 | NA | |||
633 | 20 | NOT ACTIVE | NOT ACTIVE | 3 |