Date of revision and approval

...

  • Spray the sort chamber and other appropriate internal chambers specific to the instrument with 1:100 TRIGENE and wipe clean. Spray again with 1:100 TRIGENE and allow a contact time of at least 10 minutes.
  • Wipe the areas and follow with spraying and wiping clean with 70% ethanol.
  • Spray the exposed appropriate external surfaces specific to the instrument with 1:100 TRIGENE and wipe clean. Spray again with 1:100 TRIGENE and allow a contact time of at least 10 minutes.
  • Wipe the areas sprayed and follow with spraying and wiping clean with 70% ethanol.
  • No further biological samples are to be run on the instrument until maintenance is carried out.

Emergency procedures

Emergency procedure folder can be found here in the sample prep lab:

Image Added

All emergencies need to be reported to the emergency contacts listed above. Specific chemical exposure procedures are below.

...