Attention: Confluence is not suitable for the storage of highly confidential data. Please ensure that any data classified as Highly Protected is stored using a more secure platform.
If you have any questions, please refer to the University's data classification guide or contact ict.askcyber@sydney.edu.au

Skip to end of banner
Go to start of banner

New service - LEGENDplex Th Multiplex Human Cytokine Assay

Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 22 Next »

Starting in mid July, we are going to provide the cytokine multiplex assay as a new service. This assay is validated for serum, plasma and culture supernatant.

With as little as 12.5 µl serum/plasma, or 25 µl for tissue culture samples per well, this panel allows simultaneous quantification of 13 human cytokines:















Details of the kit can be found here:

https://www.biolegend.com/en-us/products/legendplex-human-th-panel-13-plex-filter-plate-15791

Brief protocol:

  1. Thaw samples on ice. Centrifuge @ 12000g for 5min.
  2. Dilute 50 ul of serum samples 1:1 with assay buffer. Cell culture supernatant samples can be tested without dilutions but may need a preliminary experiment to determine the appropriate dilution factor.
  3. Prepare 1:4 serial dilution of standards.
  4. Pre-wet filter plate by dispensing 100 ul of wash buffer using the plate washer. Vacuum.
  5. Add 25 ul of assay buffer to each well. When pipetting aim for the side of the wells and avoid touching the membrane. This can help preventing leakages.
  6. Add 50 ul of diluted sample/standards & 25 ul of mixed beads to each well.
  7. Seal plate and shake at 500 rpm for 2 hours RT.
  8. Wash with dispensing 200 ul of wash buffer using the plate washer and vacuum.
  9. Add 25 ul of detection antibodies to each well. Seal plate and shake at 500 rpm for 1 hour RT.
  10. Add 25 ul of SA-PE to each well. Seal plate and shake at 500 rpm for 30 min RT.
  11. Wash with dispensing 200 ul of wash buffer using the plate washer and vacuum.
  12. Add 150 ul of wash buffer to each well. Resuspend beads on a plate shaker for 1 min.
  13. Samples are now ready to read on a cytometer.

Cost:

Each sample will cost ~$58 (subject to updates). We will be running 40 samples (including 1 assay control) plus 8 standard curve titrations in duplicate on a 96 well plate.

We'll also include a frozen aliquot of sample standard to monitor variation between plates.


To order this service, please complete an order for Service_Cytokine Assay under the Flow Cytometry tab on PPMS:

https://au.ppms.info/wrh/order/?pf=2


  • No labels