The recommended method that would work for users who have not recorded beads in prior experiments, would be to switch to the existing baseline (the one that existing application settings were saved on), run a CS&T using the previous bead lot, make a new experiment, apply existing application settings, copy the voltages, switch back to the new baseline, run a CS&T using the new bead lot, make a new experiment, copy the voltages over to the new experiment, and save new application settings. If any help is required please contact WRHflow well in advance as a mistake would propagate to other users. The method is outlined below.
*Note for CantoII & LSRII you will need to change to an existing configuration. Please contact staff if help is needed.
1) Run CS&T > Cytometer > CST > Performance check > using bead lot #70466 (old)
...