20180612 - LSRII UV laser
*update - 20180704 - 'B530_30B' PMT replaced. Instrument aligned. All specifications exceeded. New baseline recorded on the 4/7/2018 called WMI_2018_3B_6V_3U_3R_4Y.
It is recommended to note the UV laser change date for your reference.
If continuing application settings see below (otherwise make a new experiment and proceed as normal).
The current suggested workflow involves
- Seek assistance from WRHFlow staff if not 100% sure of the following procedure (some users may not have access to switch configurations).
- Switch to the instrument configuration with the baseline recorded on the 29/11/2017 labelled 'WMI_2015_3B_6V_3U_3R_4Y' from the current baseline recorded on the '04/07/2018' labelled 'WMI_2018_3B_6V_3U_3R_4Y'
- Run performance check 'CST' using bead lot #70466 on 'WMI_2015_3B_6V_3U_3R_4Y' configuration
- Once completed, there should be an error detected with the UV laser. Click OK and close the CST window.
- Create a new experiment.
- Apply application settings. Note the application settings applied should have been created between 29/11/2017 and '04/07/2018'.
- Record all voltages using a screen shot.
- Switch to the instrument configuration with the baseline recorded on the '04/07/2018' labelled 'WMI_2018_3B_6V_3U_3R_4Y'
- Create a new experiment, enter the voltages from the screen capture into the cytometer settings.
- Save these application settings as updated application settings that will now be linked with the baseline recorded on '04/07/2018' labelled 'WMI_2018_3B_6V_3U_3R_4Y'
- Proceed as normal applying the updated application settings to your experiments.
We strongly recommend that you make note of the UV laser change and the B530_30B PMT change in your experiment records, review the UV parameters and other parameters and record new compensation controls.
*update - 20180629 - R780_60A detector fixed.
...