- Spray the sort chamber and other appropriate internal chambers specific to the instrument with 1:100 TRIGENE and wipe clean. Spray again with 1:100 TRIGENE and allow a contact time of at least 10 minutes.
- Wipe the areas and follow with spraying and wiping clean with 70% ethanol.
- Spray the exposed appropriate external surfaces specific to the instrument with 1:100 TRIGENE and wipe clean. Spray again with 1:100 TRIGENE and allow a contact time of at least 10 minutes.
- Wipe the areas sprayed and follow with spraying and wiping clean with 70% ethanol.
- No further biological samples are to be run on the instrument until maintenance is carried out.